Dear customer, when receiving this product GE 5337137-8, GE 5337137-3, GE 5823599 RRX GE 6237000 -3 GP4 please refer to the drawings for correct connection.
As the core signal distribution and control module, Mega Switch features multi-port integration, high-bandwidth signal processing, and modular design. It supports parallel operation of Port A/B dual coil channels, enabling functions such as MC bias adjustment, MUX signal switching, and RF signal transceiving. System status is monitored in real-time through LED indicators and SRI communication.
The Mega Switch interfaces are categorized into four functional groups:
1.Coil Port Interfaces
2.Port A (J18, J21): Sends MC bias, preamp power, and MUX control signals to the coil; receives coil ID, TX enable signal, and 8-channel RF signal .
Port B (J17, J22): Performs the same functions as Port A for the Port B coil 1页.
3.External Module Interfaces
4.DM-Lite Interaction (J11, J5): Retrieves MC bias, control signals, and AUX power
1. Power Supply (J102-1, J102-2): Obtains DC power from the Regulator Box
5.RRX Communication Interfaces
6.Signal Transmission (J1): Sends 8-channel 16 MHz RF signals to RRX
Control Signals (J2, J107): Receives R1 control and R1 Strobe signals from RRX
Power and Control (J108): Transmits power, TNS control, and clock signals to RRX
7.Status Monitoring & Communication Interfaces
8.LED Indication (J15): Connects to coil ID LED indicators in LPCA
SRI Communication (J20): Enables data exchange with SRI External Signal Inputs (J100, J103-J106): Receive RF signals from Body Preamp, loopback signals from Exciter, noise signals from TNS antenna, LO signals, and 80 MHz clock from Exciter
J18: Send MC bias, preamp power, and MUX control signal to Port A coil, and get coil ID and TX enable signal.
J21: Get 8 - channel RF signal from Port A coil.
J17: Send MC bias, preamp power, and MUX control signal to Port B coil, and get coil ID and TX enable signal.
J22: Get 8 - channel RF signal from Port B coil.
J11: Get MC Bias from DM - Lite.
J5: Get control signal and AUX power from DM - Lite.
J102 - 1 and J102 - 2: Get DC power from Regulator Box.
J1: Send 8 - channel 16 MHz RF signal to RRX.
J2: Get R1 control signal from RRX.
J107: Get R1 Strobe signal from RRX.
J108: Send power, TNS control, and clock to RRX.
J15: To coil ID LED indicator in LPCA.
J20: Communication with SRI.
J100: Get RF signal from Body Preamp.
J103: Get Loopback signal from Exciter.
J104: Get noise signal from TNS antenna.
J105: Get LO signal from Exciter.J106: Get 80 MHz clock from Exciter.
We offer factory-level service testing, ensuring that every machine undergoes rigorous 99.9% inspection before shipment, following GE factory testing procedures, including signal transmission performance testing, signal reception performance testing, board-level diagnostics, RF performance testing, and communication link testing. In short, everything is done to ensure customer satisfaction and provide high-quality service.
Whatapp:+86 18562532689
Weichat:+86 18562532689
Agilent518@sina.com
Mr. Zhou's Studio
With Respect2026.03.20